7 - 6 Te-Stack Operating Manual, 391862, en, V2.3
7 - Preventive Maintenance
Adjustments and Replacements
7.5.2.2 Freedom EVO Version
Removing Remove the 4 screws (see arrows) and the covers.
Replacing/
Mounting
Place the two covers and remount the four screws.
7.5.3 Teaching the RoMa Position
Refer to the “Freedom EVOware Software Manual”.
7.5.4 Teaching the Stack Position
Perform the corresponding setups in the Te-Stack panel.
Refer to the “Instrument Software Manual”
Requested Tests and Settings
Refer to Section7.4 “Performing the Requested Tests and Settings”, 7-4
and to the “Instrument Software Manual”.
Te-MO Version: Perform the Te-Stack Functional Test procedure.
Freedom EVO
Version:
If each Transfer Station is equipped with two Te-Stacks perform the System Test
procedure, otherwise, perform the Transfer Station Functional Test procedure.
Fig. 7-1 Replacing acrylic glass covers (Freedom EVO version)