4-27
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements cont’d:
Zero Corrections cont’d
NOTE: The parasitic impedance of test adaptors and measurement cables differ from cable to
cable. Whenever a test adaptor or cable is changed, make sure that zero correction is performed
again.
Figure 4.4 - Terminal Connections for Zero Correction
Open Correction Short Correction
Connection
Normal Error Normal Error
BNC
Terminals
5-Terminal
Correction
Kelvin Klip™
Correction
Use Figure 4.4 as a reference when performing short-circuit and open-circuit zero corrections.