4200A-SCS Parameter Analyzer User's Manual Section 5: Configure and use a Series 700 Switching System
4200A-900-01 Rev. D/April 2020 5-15
4. Select Save.
5. Select the second connectpins action you added to the project tree.
6. Make the following connections using the pairs of TermIdStr# and Pin# text fields in the action:
▪ CVH1 – Pin 1
▪ CVL1 – Pin 2
▪ CVL1 – Pin 3
▪ CVL1 – Pin 4
When you are finished, the Key Parameters for the action look like the next figure.
Figure 67: Second connectpins connections
7. Select Save.