Tektronix TAS 475 - Figure 2-6: How Probe Compensation Affects Signals; Figure 2-7: Location of Probe Head Adjustment

Tektronix TAS 475
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Quick Start
Probe Compensated Correctly
Probe
Over Compensated
----Ir
l r l
__
Probe Under Compensated
Figure
2-6:
How
Probe
Compensation
Affects
Signals
11.
Adjust the low frequency compensation adjustment, located
in
the
probe body, for the best possible square wave. See Figure 2-7 for the
location of the low frequency adjustment.
Low-frequency
Adjustment
P6109B Probe Compensation
P6111
B Probe Compensation
Figure
2-7:
Location
of
Probe
Head
Adjustment
TAS
475
and
TAS
485
Service Manual
2-7
Scan by Zenith