Brief Performance Checks
4-8
11. Verify that the delayed time base
is
operational, confirming the following
statements are true.
The main sweep has
an
intensified zone of approximately 5.5 hori-
zontal divisions. Adjust the INTENSITY level if necessary to view the
entire intensified zone.
Rotating the General Purpose Knob clockwise moves the intensified
zone to the right on-screen. Position the intensified portion to the
center of the screen.
12. Adjust the
delayed sec/div scale to 200 Ils.
13. Set MAIN
to
Off and confirm that only the delayed sec/div scale
is
displayed and the period of the square-wave signal
is
about five horizon-
tal divisions.
14.
Set MAIN
to
On and DELAY to Off.
Verify the Trigger System
1.
Press the WAVEFORM OFF button until the readout indicates channel 1
is
the only selected channel.
2.
Press the AUTOSET button.
3.
Set the sec/div scale to 1 ms.
4.
Press the TRIGGER MENU button and set
MODE
to
Auto.
5.
Verify that the main trigger
is
operational, confirming that the following
statements are true.
Rotating the trigger LEVEL control changes the main trigger level
readout.
Rotating the trigger LEVEL control through its range triggers and
untriggers the
display. Leave the signal untriggered.
Pressing the SET LEVEL
TO
50% button triggers the signal and the
trigger
level readout indicates approximately 2.7
V.
6.
Press the HORIZONTAL MENU button and set DELAY
to
On.
7.
Set the delayed sec/div scale to 500 Ils.
8.
Verify that the delayed sweep
is
operational, confirming that
an
intensi-
fied zone appears on the main sweep.
9.
Press the TRIGGER MENU button and set
MODE
to
Trig After and
SRC
to
Ch1.