EasyManua.ls Logo

Tektronix TAS 475 - Figure 5-10: Adjusting Vertical Gain and Centering; Figure 5-11: High Frequency Step Response Test Setup; High Frequency Step Response

Tektronix TAS 475
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment Procedures
5-12
t----~
/-----,
100
••
.... .... .... .
...
.... ....
.
...
.
...
....
90
a
Lines Six Divisions
a
Apart,
Centered
a
Vertically
a
10
0%.'
....
....
....
.
...
....
.
...
.... .... ....
ClEAR
MENU
VLJ
l~-"'"
@
La
0 a 0
0)
Figure 5-10: Adjusting Vertical Gain and Centering
15.
Select Done when you have completed the adjustments.
16.
Select Done again to exit the routine.
High Frequency Step Response
Equipment Required: One pulse generator (item 10), one precision coaxial
cable
(item
5),
one
2X
attenuator (item
1),
and one
50
Q termination (item
3).
Adjustment Locations: This procedure requires adjustments to the Display
Driver board and the Vertical Termination hybrid. See Figures 5-14 and 5-15
on pages 5-18 and 5-19 for the
location of the Display Driver board adjust-
ments.
See Figure 5-13 on page 5-17 for the location of the Vertical
Termination hybrid adjustment.
Prerequisites: Factory
Vertical
Cal
adjustment procedure.
1.
Display channel
1,
turning all others off.
2.
Connect the positive fast rise pulse output of the pulse generator to the
CH 1 input
as
shown
in
Figure 5-11.
p
ulse
Generator
=-0
a a
a a
"ii.
a
~.
'0'
rO
a
. 0
a
. a
ro
lo
a
·0
10
E3
@
a
a a
a a a a a
....,.0
a
0
0
.:r
'-"
'-" '-"
'-"
EI
fiiI
=(j)=(j)=(j)~
/q
500
Te~;n'"M
~~
Precision Cable
) 2XAtte
nuator
Figure 5-11: High Frequency Step Response Test Setup
Adjustment
Scan by Zenith

Table of Contents

Related product manuals