VERTICAL Controls
THS710A, THS720A, THS730A & THS720P User Manual
3–69
Channel 1 or Channel 2 Vertical Menu
The vertical menu contains the following items when channel 1 or
channel 2 is the selected waveform.
S
PE VE
L
MENU
Coupling DC
AC
GND
Invert Invert Off
Invert On
Bandwidth Full Bandwidth
20 MHz
Position —
P
o
yp
Current Probe Set conversion
factor
Voltage Probe Set probe
attenuation
Key Points
GND Coupling. Use GND coupling to display a zero-volt waveform.
When you use GND coupling, the input BNC connector is
disconnected from internal circuits. Internally, the channel input and
its reference are connected to create a zero-volt reference level.
VOLTS/DIV Rocker. Use the VOLTS/DIV rocker to change the vertical
sensitivity when acquisition is running. When acquisition is stopped,
the rocker scales the waveform vertically.