Appendix D: Performance Verification
D–4
THS710A, THS720A, THS730A & THS720P User Manual
Performance Verification Procedures
Before beginning these procedures, two conditions must first be met:
H The TekScope instrument must have been operating continuously
for ten minutes in an 18° C to 28° C ambient environment with a
relative humidity of less than 60%.
H You must perform the Compensate Signal Path operation
described on page 2–11. If the operating temperature changes by
more than 5° C, you must perform the Compensate Signal Path
operation again.
The time required to complete the entire procedure is approximately
one hour.
WARNING. Some procedures use hazardous voltages. To prevent
electrical shock, always set voltage source outputs to 0 V before
making or changing any interconnections.