EasyManuals Logo

Tektronix TekScope THS710A User Manual

Tektronix TekScope THS710A
231 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
Appendix D: Performance Verification
D–4
THS710A, THS720A, THS730A & THS720P User Manual
Performance Verification Procedures
Before beginning these procedures, two conditions must first be met:
H The TekScope instrument must have been operating continuously
for ten minutes in an 18° C to 28° C ambient environment with a
relative humidity of less than 60%.
H You must perform the Compensate Signal Path operation
described on page 2–11. If the operating temperature changes by
more than 5° C, you must perform the Compensate Signal Path
operation again.
The time required to complete the entire procedure is approximately
one hour.
WARNING. Some procedures use hazardous voltages. To prevent
electrical shock, always set voltage source outputs to 0 V before
making or changing any interconnections.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TekScope THS710A and is the answer not in the manual?

Tektronix TekScope THS710A Specifications

General IconGeneral
BrandTektronix
ModelTekScope THS710A
CategoryTest Equipment
LanguageEnglish

Related product manuals