Appendix D: Performance Verification
THS710A, THS720A, THS730A & THS720P User Manual
D–5
Self Test
This procedure uses internal routines to verify that the TekScope
instrument functions and passes its internal self tests. No test
equipment or hookups are required. Start the self test with the
following setup:
S
PE
L
System Diag
Loop Once
Execute
OK Run Test
A dialog box displays the result when the self test completes. Press
the CLEAR MENU button to continue operation.