Appendix A: Specifications
This appendix contains the oscilloscope, DMM, and general
specifications for the THS710A, THS720A, THS730A, and
THS720P TekScope instruments. All specifications are guaranteed
unless noted as “typical.” Typical specifications are provided for
your convenience but are not guaranteed. Specifications that are
marked with the n symbol are checked in Appendix D: Performance
Verification.
All specifications apply to all the TekScope instruments unless noted
otherwise. All specifications assume horizontal MAG is off, unless
noted otherwise. To meet specifications, two conditions must first be
met:
H The TekScope instrument must have been operating continuously
for ten minutes within the operating temperature range specified.
H You must perform the Compensate Signal Path operation
described on page 2–11. If the operating temperature changes by
more than 5° C, you must perform the Compensate Signal Path
operation again.
Oscilloscope Specifications
Acquisition
Acquisition Modes Sample (Normal), Peak detect, Envelope, and Average
Acquisition Rate,
typical
Up to 25 waveforms per second (2 channels, sample acquisition
mode, MAG on, no measurements)
Singl
S
u
nc
Acquisition Mode Acquisition Stops After
Sample, Peak Detect Single acquisition, one or two
channels simultaneously
Average, Envelope N acquisitions, one or two
channels simultaneously, N is
settable from 2 to 256 or ∞