Appendix D: Performance Verification
D–6
THS710A, THS720A, THS730A & THS720P User Manual
Check DC Measurement Accuracy
1. Set the DC voltage source output level to 0 V.
2. Set up the TekScope instrument using the following steps:
S
CO
PE S
A
VE/
RECALL
Recall Saved
Setup
Recall Factory
Setup
OK Recall
Factory
CH 1
VERTICAL
MENU
Probe Type Voltage Probe Set to 1X
ACQUIRE Acquire Mode Average Set to 16
ME
A
S
Select
Measrmnt
Mean*
OK Select
Measrmnt
* You may need to press Select Page to display this selection.
3. Connect the TekScope instrument to the DC voltage source as
shown below.
+
DC voltage
source
Ch 1
Dual banana
to bnc adapter
BNC cable