DMM Tests Low Limit Test Result High Limit
DC
Voltag
e
Accuracy
400 mV
range,
60 mV input
59.2 mV 60.8 mV
400 mV
range,
360 mV input
357.7 mV 362.3 mV
4 V range 3.577 V 3.623 V
40 V range 35.77 V 36.23 V
400 V range 357.7 V 362.3 V
880 V range 783 V 801 V
AC
Voltag
e
Ar
400 mV range 352.3 mV 367.7 mV
A
ccu
r
acy
4 V range,
600 mV input
0.583 V 0.617 V
4 V range,
3.6 V input
3.523 V 3.677 V
40 V range 35.23 V 36.77 V
400 V range 352.3 V 367.7 V
640 V range 559 V 593 V
Res
i
s
tanc
e
Ar
400 range 358.0 362.0
A
ccu
r
acy
4 k range
3.580 k 3.620 k
40 k range,
6 k input
5.95 k 6.05 k
40 k range,
36 k input
35.80 k 36.20 k
400 k range 358.0 k 362.0 k
4 M range 3.580 M 3.620 M
40 M range 35.23 M 36.77 M