Appendix D: Performance Verification
D–10
THS710A, THS720A, THS730A & THS720P User Manual
Check Channel 2 Bandwidth
1. First check the channel 1 bandwidth using the previous test.
Then, perform these additional steps to check the channel 2
bandwidth:
S
CO
PE
CH 1
WAVE-
FORM
OFF
CH 2
TRIGGER Trigger Source Ch2
ME
A
S
Gating & High-
Low Setup
High-Low
Method
Min/Max
Select
Measrmnt
Pk-Pk*
OK Select
Measrmnt
* You may need to press Select Page to display this selection.
2. Connect the TekScope instrument to the leveled sine wave
generator as shown below.
Leveled
sine wave
generator
Ch 2
Output
50 feedthrough
terminator
BNC cable