General-Purpose Application Examples
THS710A, THS720A, THS730A & THS720P User Manual
2–33
Setup to Trigger on the Odd Field
S
PE
VERTICAL
MENU
Probe Type Voltage
Probe
Set to 1X
AUTO
RANGE
— — —
DISPLAY Display Style Dot
Accumulate
Set to 100 ms
E
Trigger Type Video
ME
Trigger On Odd Field
Video Class NTSC
Adjust the SEC/DIV rocker to 2 ms/div to display the odd field
across about eight divisions. The Dot Accumulate display style
simulates an analog oscilloscope display of the video signal.
Going Further
Use the following setup to trigger on a specific video line.
S
PE
E
Trigger Type Video —
ME
Trigger On Odd Field Set line
number
Adjust the SEC/DIV rocker to 10 s/div to display the line across
about six divisions.