ACQUIRE
3–4
THS710A, THS720A, THS730A & THS720P User Manual
Samples acquired in four
acquisition intervals
Displayed record
points
Acquisition
mode
Sample mode acquires one sample in each interval.
Peak Detect
Interval 1 2 3 4 Interval 1 2 3 4
Peak Detect mode uses the lowest and highest samples from two consecutive intervals.
Sample
Sample. Use Sample acquisition mode for the fastest acquisition at
any SEC/DIV setting. Sample mode is the default mode.
Peak Detect. Use Peak Detect acquisition mode to limit the possibility
of aliasing. Also, use Peak Detect for glitch detection. You can see
glitches as narrow as a 8 ns.
Peak Detect is only functional for sample rates up to 25 MS/s. For
50 MS/s and faster sample rates, the TekScope instrument switches