Appendix A: Specifications
Ta ble 6: Oscilloscope Inputs (cont.)
Characteristic Description
Scan rates
14–22 kHz
19–33 kHz
31–65 kHz
Pulse Width Trigger
Screen update On Trigger, Single Shot
Trigger conditions
<T, >T, =T (±10%), ≠T(±10%)
Source Ch1
Polarity Positive or negative pulse
Sensitivity
same as edge trigger
Level same as edge trigger
Pulse time adjustment range
0.01 div. to 655 div. with a minimum of 300 ns (<T) or
500 ns (=T, ≠T), a maximum of 10 s, and a resolution of
0.01 div. with a minimum of 50 ns
Continuous Auto Set
Auto-ranging attenuators and time base, automatic AutoRange triggering with automatic source selection.
Modes
Normal 15 Hz to maximum bandwidth
Low frequency
1 Hz to maximum bandwidth
Minimum amplitude Ch1, Ch2, Ch3, Ch4
DC to 1 MHz
10 mV
1 MHz to maximum bandwidth 20 mV
Minimum trigger amplitude 55 mV rms
Automatic Capturing Screens
Capacity
100 screens
For viewing screens, see Data Log function.
THS3000 Series Oscilloscopes User Manual 85