IM-LYOQUEST-EN-2101 45 / 58
11.3. Refrigerating system maintenance
The refrigerating system requires minimal maintenance. If due to accidents or breakages oil
or refrigerant needs to be added, it will be necessary to get in contact with Azbil Telstar
Technologies Customer Service.
To assure the proper refrigeration of the system, all the grids must be at a minimum
distance of 20cm away from any other equipment and from the wall.
It is recommended to perform a periodic cleaning of the dust accumulated in the condenser
batteries of the refrigerating system. To proceed, the lateral doors of the freeze dryer must
be opened to clean the battery by means of a small paintbrush, and if it is possible by means
of compressed air. This should be done particularly before summer.
When the switch is off, the compressor continues working for the time set by the collecting
timer (2 minutes) in order to inhale the gas of the evaporator. If the time has passed and
the compressor has not stopped, the equipment will have to be checked, for it’s not
advisable that the compressor works without gas circulation.
Refrigerant loads
LyoQuest -55 LyoQuest -85
R-23 ----- 95 g
R-507 420 g 390 g
LyoQuest Arctic -85
R-170 70 g
R-290 125 g
11.4. Maintenance table
The frequencies for each operation, particularly the oil change in the vacuum pump, that are
shown in the following Maintenance Table, are suggested values for a normal operation of
the freeze-dryer. Extremely severe working conditions or the use of aggressive fluids can
reduce the intervals mentioned in the Table.
OPERATION PERIODICITY
VACUUM UNIT
- Oil condition control Daily
- Change of oil filter 1 year
- Oil level control Daily
- Visual inspection of oil leaks Daily
- First oil change After the first 150 hours of service
- Oil change Minimum every 500 hours of service
- Pump body cleaning 1 year
- Check the performance of the vacuum unit,
particularly the standard time for reaching
vacuum throughout the unit with cold condenser
P
atm
to 1mbar < 5min
6 months
REFRIGERATING UNIT
- Condenser panel cleaning 6 months
- Check the performance of the refrigerating
system, verifying the normal times to cool the
condenser.
LyoQuest -55: +20 to -30 <10 min
LyoQuest -85: +20 to -40 <5 min
LyoQuest Arctic -85: +20 to -40 <5 min
6 months
FREEZE DRYER
- Joint cleaning and greasing 6 months