EasyManua.ls Logo

TEM Aligna 4D - Page 71

Default Icon
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Aligna
®
4D User Manual
71 / 84
10.4.3.13 Manual edition of the Script "Default"
You can also edit all relevant parameters manually.
NOTE: Most of the numerical parameters inside the µC are handled as integer values.
Even if the display in Kangoo might be in decimal values (displayed Value "PSD A gain" is
12.34 [%]) the command to the µC is given as integer ("PsdAgain= 1234"). (The scaling
factor between the displayed Kangoo device value (12.34) and the sent µC command val-
ue (1234) is defined in the device's "gain" property, "0.01" in this case.)
If you want to see the correct syntax (including the correct variable name and the number of
digits or scaling resp.), switch on the COM terminal (menu item "Communication / display
COM window (F6)"), scroll the wanted parameter, e.g., with the mouse wheel and watch the
communication. ("\PsdAgain= 1234". The backlash "\" suppresses the µC echo. It should NOT
be used in the script commands ("PsdAgain= 1234"), because you want the µC to echo these
commands. Thus, Kangoo can update its displayed devices, receiving these echoes during
boot process.
When you save the modified script file, Kangoo observes this file, and when it is changed,
Kangoo will ask you, whether you want to flash it to the µC's EEPROM now.
Now you are able to set any changed value as default value and you can restore whenever
you want.