12 Thorlabs Beam 4.0
2.4.4
Trigger Input
An electrical TTL trigger input is provided for synchronizing the Camera Beam
Profiler's global shutter to Pulsed Laser Sources . The appropriate BNC
connector is situated on the left side of the instrument body, see Connectors .
The following table summarizes important electrical data for a compatible signal
source.
Parameter
Save static voltage level
Max. low level
Min. high level
Input impedance
Min. pulse width
Min. slew rate
Applicable safety standards
Attention
Be sure to enter only a TTL compatible signal to the BNC jack not exceeding
the range (-0.5 V … +6.5 V). Higher or lower voltages may damage the Beam
Profiler input!
Note
Use shielded coaxial cables for connecting the trigger source only. Do not connect a
trigger cable without a signal source because it may capture disturbing trigger
pulses.