151Appendix
7.4.2
Wavelength Response
1. Wavelength Dependent Sensitivity of BC106 Models
Image sensors of both BC106 models show a wavelength depending response. The
following diagrams show its relative response. Whereas model BC106-VIS offers
about double sensitive in the visible wavelength range, it suffers from nearly zero
sensitivity in the UV range below 300 nm.
As shown here in detail the sensitivity of model BC106-UV considerable exceeds
model BC106-VIS below 300 nm.
Both response diagrams are measured without any ND Filter.