Page #293 background imageLoading...
Page #293 background image
3-237
l) OA Connection Setup
(1) Patient ID (US/Opt)
Sets how to handle the measurement data according to the ID in the ultrasonic (US) mode
and light interference (Opt) mode.
Common ID: The same ID is used in both of the ultrasonic (US) mode (axial length
measurement, corneal thickness measurement) and light interference (Opt)
mode.
The ID and measurement data are not cleared even
when the axial length measurement and corneal
thickness measurement are switched in the ultrasonic
(US) mode.
For a new patient, the ID and measurement data in all
modes are cleared.
Individual ID: Separate IDs can be used in the ultrasonic (US) mode and light interference
(Opt) mode.
The ID is cleared when the axial length measurement
and corneal thickness measurement are switched in
the ultrasonic (US) mode.
3-159
(1)
(3)
(2)