3-196
3.11.3 Transferring measurement data (Axial length measurement
display/Corneal thickness measurement display)
■
The network support system “TOMEY Link” (optional)
or the inspection data receiving software “DATA
Transfer” (included in the package) is required for data
communication.
■
Refer to the corresponding operation manual for the
TOMEY Link and DATA Transfer settings.
■
Connection setting is required for connecting with
TOMEY Link and DATA Transfer. Refer to “3.12
.
2 d)
TOMEY Link setup” to make the setting.
■
Data cannot be sent in some configurations if the ID
number is not entered. Refer to “3.12.2 g) Bluetooth
setup” for details.
■
It is recommended to send data after entering the ID
number.
The data can be sent without ID numbers entered;
however, the corresponding patient may not be traced
from the inspection file output from DATA Transfer.
When the data was sent without entering the ID number,
check the content of the inspection file immediately
after sending the data, and move and save the file in an
appropriate location. In addition, when data was sent
to TOMEY Link, all the data without ID number are
regarded as the data of the same patient so that they
must be handled with great care to avoid mishandling.
Refer to “3.3.6 i) Function to transfer (EXPORT) patient data” and “3.3.6
l) Group export function” for the data transfer function in the B-mode
image diagnosis and the axial length measurement auxiliary function.