5-9
■
5.5.2 Fuses
■
Disconnect the power cord from the power socket
before replacing. Otherwise you may get an electric
shock, resulting in death or serious injuries.
■
Use fuses specifically designed for this
instrument.(T2AL, 250V)
■
When the instrument does not work correctly after
fuses have been replaced, there may be other causes
of the problem. Turn off the instrument immediately
and contact our local distributor.
1) Turn power off and disconnect the power plug from the outlet.
2) Disconnect the power plug from the power terminal.
3) Insert the tip of a flat screwdriver into the groove on the top of the fuse
holder under the power terminal, release the lock, and remove the holder
case.
4) Remove the blown fuse from the fuse holder case and install a new fuse.
5) Return the holder case to the fuse holder in the reverse order of removal.