3-166
3.8.4 Checking the measurement data
a) Intraocular pressure correction
It is possible to perform intraocular pressure correction after measurement.
(Refer to “3.8.2 f) Intraocular pressure correction function”)
b) Deleting part of measurement data
1) Touch a memory number (1) to select the data to be deleted.
2) Touch the “Delete” button (2) to delete the selected data, and the average
and the standard deviation of corneal thickness are recalculated.
3) If you have deleted data mistakenly or want to cancel the deletion, select
that data number and you will see that the “Delete” button (2) will change to
“Return” button (3). Touch the “Return” button to cancel the deletion. When
new measurement data is captured, however, the deleted data cannot be
restored by pressing the “Return” button.
3-115
(Fig. 1)
(2)
(1)