4-1
4. TECHNICAL INFORMATION
4.1 IOL power calculation formula
4.1.1 SRK-II formula
1. Implanted IOL power (D) for emmentropization
LKAPemme 5.29.01
=
2. Implanted IOL power (D) for ametropia
CRREFPP emmeamet
=
3. Predicted refractive power after surgery (D)
(
)
CRPPREF emmeiol /
=
Where:
L<20.0 A1 = A +3
20.0 L < 21.0 A1 = A +2
21.0 L < 22.0 A1 = A +1
22.0 L < 24.5 A1 = A
24.5 L A1 = A – 0.5
emmeP
14 CR = 1.00
emmeP
> 14 CR = 1.25
A : A-Constant
K : Average corneal refractive power (D)
= (K1 + K2)/2
r : Radius of corneal curvature (mm)
= {(KI - 1) × 1000}/K
KI : Cornea equivalent refractive index
L : Axial length (mm)
P : Implanted IOL power (D)
REF : Desired refractive power after surgery