EasyManuals Logo

Tomey UD-8000 User Manual

Tomey UD-8000
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #343 background imageLoading...
Page #343 background image
6-9
(indicated by a pulldown button on the right side
of the thumbnail) are all matched. If needed,
display thumbnail of desired data by changing
the patient ID, patient's eye, or probe type.
z Four images are not displayed with the four-images display function.
Four or more images are not displayed on the
video monitor or static image monitor.
Display four or more images on the video
monitor or static image monitor if needed.
z Display of the video does not follow quick motions of the measured eye.
The smoothing function is in use.
Verify that the smoothing function is effective. If
it is effective, disable it and verify its result.
z Operation of NEAR and FAR volume cannot be performed.
Operation is in the freeze mode.
You can adjust near gain or far gain only in the
real time mode. Terminate the freeze mode.
However, this will erase video stored.
z A probe error is displayed during starting operation or switching to the B-mode
image diagnostic function.
The B-mode probe is not connected correctly.
Check that the B-mode probe is firmly
connected.
z No comment is displayed on images.
The display setting of comments is “hidden.”
Turn on the display of comments as described
in “3.9.2 i) Display setting within images.”