3-12
i) Connecting a personal computer
When operating this instrument connected to a
computer for Data
Transfer, only use devices that
satisfy IEC60601-1 or equivalent safety requirements
or that conform to IEC60950 and whose power source
is insulated with an isolation transformer, in order to
satisfy the relevant safety requirements for medical
equipment.
The network support system “TOMEY Link” (optional)
or the inspection data receiving software “DATA
Transfer” (included in the package) is required for data
communication with the Machine.
Refer to the corresponding operation manual for
installation, settings, and operation of TOMEY Link and
DATA Transfer.
Appropriate connection settings are required for
connection with TOMEY Link and DATA Transfer.
Refer
to “3.10.2 d) Tomey Link connection settings” to make
settings.
Be sure to make network settings under the consent
of your network administrator.
There are two methods to connect with a personal computer: LAN
connection and USB connection.