User's Manual
750 50 0042 03 12
6. Slide the measuring insert into the holder and lock
it, using the knob. Take care to position the
eccentric probes correctly.
7. Release the transport safety screw.
8. Check / adjust the floating probe suspension
balancing (§ 14.2).
9. If the instrument does not switch on or in case of
low battery level, proceed to a full loading of the
batteries (plug in the charging unit to the
instrument).
Only use the charging unit supplied with the instrument to load the batteries. Empty
batteries are fully loaded in about 12 hours. It is not necessary to wait until the
battery pack is fully charged. The instrument becomes operational immediately after
having connected the charging unit.
It is not dangerous to leave the charging unit connected. Permanently plugged-in
instruments automatically switch to trickle charge mode.
The batteries are of Lithium-ion type. An excessive room temperature can affect the
battery capacity and therefore the instrument autonomy. It is not recommended to
load the battery in case of room temperature superior to 40C°.
Incorrect loading of the battery can result in a decrease of capacity, overheating, or
even an explosion and cause important damages.
The battery can be loaded and unloaded 300 times before a significant reduction of
their capacity. The number of loading cycles and the autonomy can vary according
to use and operating conditions.