User's Manual
750 50 0042 03 72
14.2 Adjustment of Floating Probe Suspension Balancing
In order to guarantee a constant measuring force in both
directions (upwards and downwards), it is necessary to
adjust the balancing of the floating probe suspension
according to the probe in use.
The adjustment Wheel (V7) or lever (V9) of the floating
suspension (1) allows compensating the weight of the
probe in use.
14.2.1 With Lever Dynamometer
A lever dynamometer is the most exact method of
checking the floating suspension balancing.
1. Press the key on the dynamometer until the acoustic
signal is heard and read the displayed force.
2. Now, repeat the operation in the other measurement
direction and compare the results.
3. Adjust the balancing if required.
14.2.2 With Function Integrated in the Instrument
The instrument offers a balancing help function.
1. Enter the configuration menu and scroll to balancing
adjustment.
2. Pull the dual carriage downwards until stop and
release so that it returns to its balance position. The
corresponding field must display
. Else, adjust the
balancing.
3. Do the same by pushing the dual carriage upwards.