EasyManuals Logo

Troxler EGauge Combo 4540 User Manual

Troxler EGauge Combo 4540
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
NOTE
The source rod should always be in the SAFE position
when the gauge is not in use.
Instructions for Taking a Standard Count
The standard count consists of 2 separate 2-minute counts.
The first count occurs with the handle in SAFE position.
To ensure that the handle is seated in the Safe position
properly, lightly press down on it (without engaging the
trigger) just before starting the standard count.
During the second part the handle is in background position.
1. Prepare the test site as instructed on page 42.
2. Take a standard count by pressing the STD key.
3. Press YES to start the standard count:
4. Make sure the gauge is on the material to be measured and
ensure that the source rod in the SAFE position.
NOTE
If not in correct position this warning displays:
If you see this message, place the source rod in the
SAFE position and press <ENTER>.
With Gauge on
Test Site & Source
Rod in BG Pos.
Press <START>
Standard Count
DS= xxxx DSB= xxxx
Take New Count?
<YES> or <NO>
Source Rod
NOT in Safe
Position
Press <ENTER>
4-6

Table of Contents

Other manuals for Troxler EGauge Combo 4540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler EGauge Combo 4540 and is the answer not in the manual?

Troxler EGauge Combo 4540 Specifications

General IconGeneral
BrandTroxler
ModelEGauge Combo 4540
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals