Model 4540
512
To select a menu option, press the corresponding numeric key
or press ESC to return to the Setup menu.
Taking a Drift Test
To take a new drift test, press 1 at the Drift Test menu. Place
the gauge on a level surface with the source rod in the SAFE
(shielded) position, and then press ENTER/START.
NOTE
If it has been less than three hours or more than eight
hours since the last stat test, the gauge displays an
error message. Press ENTER/START to return to the
Setup menu.
The gauge displays the progress of the drift test (five 4-minute
counts). Upon completion of the counts, the gauge displays the
results of the density drift test, including an indication of
whether the test passed or failed:
Press ENTER/START to view the results of the moisture drift
test. Follow the prompts to scroll through the results.
Reviewing the Drift Test
To review the results of the drift stat test, press 2 at the Drift
Test menu. The gauge displays the results of the density drift
test.
Density Drift Test
Avg. Counts: ####
R: #.### (Pass)
<ENTER> for Moist.
5-14

Table of Contents

Other manuals for Troxler EGauge Combo 4540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler EGauge Combo 4540 and is the answer not in the manual?

Troxler EGauge Combo 4540 Specifications

General IconGeneral
BrandTroxler
ModelEGauge Combo 4540
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals