C–10 Series 3080 Electrostatic Classifiers
SIS Return: read serial number of all impactor
example:
0457 cm; SN: 320
0508 cm; SN: 120
0710 cm; SN: 173
Set Offset Bit Value of Flow Sensor
SOVx (Set Offset Values of pressure transducer and flowmeter)
SOVx Set: measure offset bit values the following sensors
and stores values in EEPROM:
x = B: Bypass Pressure Transducer
x = I: Impactor Pressure Transducer
x = L: Flowmeter
SOV Return: offset of Bypass Pressure Transducer (in bit),
offset of Impactor Pressure Transducer (in bit),
offset of Flowmeter (in bit)
Set Sheath Flow Calibration Parameter A
(Polynomial Flow Calibration Algorithm)
SCSAx (Set Calibration Sheath Parameter A)
Set sheath flow calibration parameter A:
SCSA1.234567e-12 Set: sheath flow calibration parameter A
SCSA Return: sheath flow calibration parameter A
Set Sheath Flow Calibration Parameter B
(Polynomial Flow Calibration Algorithm)
SCSBx (Set Calibration Sheath Parameter B)
Set bypass flow calibration parameter B:
SCSB1.234567e-12 Set: sheath flow calibration parameter B
SCSB Return: sheath flow calibration parameter B
Set Calibration Bypass Parameter A
(Polynomial Flow Calibration Algorithm)
SCBAx (Set Calibration Bypass A Parameter)
Set sheath flow calibration parameter:
SCBA1.234567e-12 Set: bypass flow calibration parameter
SCBA Return: bypass flow calibration parameter