Page 93
Message / Symbol Description
paramerter may need to
be set.
This message is displayed if I/O
card is faulty (both while cycle is
running or not).
Turn the autoclave off
and on again. If the
problem persists, call for
service.
"I/O card is
not
connected"
This message is displayed if I/O
card is disconnected (both while
cycle is running or not).
Turn the autoclave off
and on again. If the
problem persists, call for
service.
This message is displayed if the
temperature drops for more than 1
second below the sterilization
temperature during sterilization
cycle.
This message is displayed if the
temperature raises 4°C above
sterilization temperature during the
sterilization stage for 2 seconds
during sterilization cycle.
This message is displayed if the
system cannot reach the required
temperature, in the chamber,
within 10 minutes.
This message is displayed if the
system cannot reach the required
temperature, in the chamber,
within the preset time.
Verify that the autoclave
is not overloaded.
This message is displayed if
Chamber Pressure drops below
the sterilization pressure for 2
seconds during the sterilization
stage.
This message is displayed if
Chamber Pressure raises 29 kPa
above sterilization pressure for 2
seconds during the sterilization
stage.
This message is displayed if the
system cannot reach atmospheric