Page 94
Problem/
Error
Message
Message / Symbol Description
Corrective Action
(Ending)"
pressure ± 5kPa during the ending
stage.
"High
Pressure
(Exhaust)"
This message is displayed if the
system cannot reach preset
pressure within 10 minutes from
the beginning of the exhaust stage.
Perform a new cycle.
"High
Pressure
(Dry)"
This message is displayed if the
pressure in chamber exceeds
atmospheric pressure by more
than 10kPa at the beginning of the
dry stage.
Perform a new cycle.
"Pressure
Time Error"
This message is displayed if the
system cannot reach the required
pressure conditions in the
chamber, after preset time, during
the air removal stage.
Verify that the autoclave
is not overloaded.
"RTC Error -
Please Set
Current Date
and Time"
This message is displayed in
order to set the date and the time.
Set Current Date and
Time.
If the problem persists,
call the technician.
"Time Error"
This message is displayed if the
real time clock is faulty.
Call for service.
"Door is open
(During the
cycle)"
This message is displayed when
the door is open: During the cycle.
Close the door to perform
a new cycle.
"Canceled By
User"
This message is displayed after
the START/STOP key is pressed
and cycle aborted.
Wait until "cycle failed
canceled by user" or
"cycle end canceled by
user" is displayed.
Perform a new cycle.
"Cycle Failed"
This message and symbol are
displayed if an error occurs before
sterilization cycle is completed.
Perform a new cycle.
"Test Fainled"
This message and symbol are
displayed if an error occurs before
test cycle is completed.
Perform a new test.
"Air Error"
This message is displayed at the
end of the cycle if the autoclave
Wait until the autoclave
reaches the atmospheric