EasyManua.ls Logo

UNI-T UTG9000T Series

UNI-T UTG9000T Series
162 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UTG9000T Series Function Abitrary Waveform Generator
20
Figure 2-4 Frequency Settings
2.3.2 Output Amplitude Settings
The instrument output of a sine wave amplitude is 100mV peak value (default setting) when activating
the instrument.
The step to set the amplitude to 300mVpp:
1Tap Amplitude tab,pop-out the virtual numeric keyboard to enter 300mVpp
2Tap word Amplitude to step through unit of VppVrmsdBm
Note: dBm setting only enable when Load is no HighZ mode
Figure 2-5 Amplitude Settings
2.3.3 DC Offset Voltage Settings
The instrument output DC offset voltage of a sine wave amplitude is 0V (default setting) when activating
the instrument.
The step to set the DC offset voltage to-150mV:
1Tap Continue tab to select Sine
2Tap Offset tab, pop-out the virtual numeric keyboard to enter -150mV
3Tap word Offset, Amplitude and Offset tab becomes High (maximum)/Low (minimum) leve. This
method is convenient to set the signal limits of digital applications

Table of Contents

Related product manuals