EasyManuals Logo
Home>YOKOGAWA>Test Equipment>CA150

YOKOGAWA CA150 User Manual

YOKOGAWA CA150
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
Specications
11-1
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
11. Specifications
100 mVDCV
source
DCmA
source
mA SINK
Resistance
source
RTD output
TC output
1 V
10 V
30 V
20 mA
20 mA
SINK
500 Ω
5 kΩ
50 kΩ
PT 100
JPT 100
K
E
J
T
N
L
U
0 to ±110.000 mV
0 to ±1.10000 V
0 to ±11.0000 V
0 to ±30.00 V
0 to +22.000 mA
0 to -22.000 mA
0 to 550.00 Ω
0 to 5.5000 kΩ
0 to 55.000 kΩ
- 200.0 to 850.0 °C
- 200.0 to 500.0 °C
-200.0 to 1372.0 °C
-200.0 to 1000.0 °C
-200.0 to 1200.0 °C
-200.0 to 400.0 °C
-200.0 to 1300.0 °C
-200.0 to 900.0 °C
-200.0 to 400.0 °C
±(0.02% +10 µV)
±(0.02% +0.05 mV)
±(0.02% +0.5 mV)
±(0.02% +10 mV)
±(0.025% +3 µA)
±(0.025% +6 µA)
±(0.02% +0.1 Ω)
±(0.05% +1.5 Ω)
±(0.1% +50 Ω)
±(0.025% +0.3 °C)
±(0.02% +0.5 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.8 °C)
±(0.02% +0.4 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.6 °C)
±(0.02% +0.4 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.7 °C)
±(0.02% +0.5 °C)
However, -200.0 to -100.0
°C
±(0.02% +0.8 °C)
±(0.02% +0.5 °C)
However, -200.0 to 0 °C
±(0.02% +1.0 °C)
±(0.02% +0.5 °C)
±(0.02% +0.5 °C)
However, -200.0 to 0 °C
±(0.02% +0.7 °C)
Output resistance:
Approx. 6.5 Ω
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Maximum load: 24 V
External power supply:
5 to 28 V
Excitation current:
1 to 5 mA*2 or maximum
output: 2V
Excitation current:
0.1 to 0.5 mA or
maximum output: 2V
Excitation current:
0.01 to 0.1 mA or
maximum output: 2V
Excitation current:
1 to 5 mA*2
*3 TC source accuracy
does not include RJ
sensor accuracy.
RJ Sensor Specification
Measurement range:
-10 to 50 °C
Accuracy
(when combined with
main unit)
18 to 28 °C: ±0.5 °C
Other than 18 to 28°C:
±1 °C
1 µV
Range Source Range Accuracy
Remarks
Setting
Resolution
10 µV
0.1 mV
10 mV
1 µA
1 µA
0.01 Ω
0.1 Ω
1 Ω
0.1 °C
0.1 °C
*1
*1
*3
Source
Accuracy = ± (% of setting + µV, mV, µA, , °C, CPM, Hz, kHz) at 23±5°C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA CA150 and is the answer not in the manual?

YOKOGAWA CA150 Specifications

General IconGeneral
BrandYOKOGAWA
ModelCA150
CategoryTest Equipment
LanguageEnglish

Related product manuals