Specications
11-1
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
11. Specifications
100 mVDCV
source
DCmA
source
mA SINK
Resistance
source
RTD output
TC output
1 V
10 V
30 V
20 mA
20 mA
SINK
500 Ω
5 kΩ
50 kΩ
PT 100
JPT 100
K
E
J
T
N
L
U
0 to ±110.000 mV
0 to ±1.10000 V
0 to ±11.0000 V
0 to ±30.00 V
0 to +22.000 mA
0 to -22.000 mA
0 to 550.00 Ω
0 to 5.5000 kΩ
0 to 55.000 kΩ
- 200.0 to 850.0 °C
- 200.0 to 500.0 °C
-200.0 to 1372.0 °C
-200.0 to 1000.0 °C
-200.0 to 1200.0 °C
-200.0 to 400.0 °C
-200.0 to 1300.0 °C
-200.0 to 900.0 °C
-200.0 to 400.0 °C
±(0.02% +10 µV)
±(0.02% +0.05 mV)
±(0.02% +0.5 mV)
±(0.02% +10 mV)
±(0.025% +3 µA)
±(0.025% +6 µA)
±(0.02% +0.1 Ω)
±(0.05% +1.5 Ω)
±(0.1% +50 Ω)
±(0.025% +0.3 °C)
±(0.02% +0.5 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.8 °C)
±(0.02% +0.4 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.6 °C)
±(0.02% +0.4 °C)
However, -200.0 to -100.0 °C
±(0.02% +0.7 °C)
±(0.02% +0.5 °C)
However, -200.0 to -100.0
°C
±(0.02% +0.8 °C)
±(0.02% +0.5 °C)
However, -200.0 to 0 °C
±(0.02% +1.0 °C)
±(0.02% +0.5 °C)
±(0.02% +0.5 °C)
However, -200.0 to 0 °C
±(0.02% +0.7 °C)
Output resistance:
Approx. 6.5 Ω
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Maximum load: 24 V
External power supply:
5 to 28 V
Excitation current:
1 to 5 mA*2 or maximum
output: 2V
Excitation current:
0.1 to 0.5 mA or
maximum output: 2V
Excitation current:
0.01 to 0.1 mA or
maximum output: 2V
Excitation current:
1 to 5 mA*2
*3 TC source accuracy
does not include RJ
sensor accuracy.
RJ Sensor Specification
Measurement range:
-10 to 50 °C
Accuracy
(when combined with
main unit)
18 to 28 °C: ±0.5 °C
Other than 18 to 28°C:
±1 °C
1 µV
Range Source Range Accuracy
Remarks
Setting
Resolution
10 µV
0.1 mV
10 mV
1 µA
1 µA
0.01 Ω
0.1 Ω
1 Ω
0.1 °C
0.1 °C
*1
*1
*3
Source
Accuracy = ± (% of setting + µV, mV, µA, , °C, CPM, Hz, kHz) at 23±5°C

Table of Contents