Memory Function
6-3
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
6.1.1 Saving
1 Press the SAVE key when a source value or measurement value is
displayed.
MEM No. (memory number) appears.
The next number after the largest used (saved) number appears.
2 Use the ▲ ▼ key to change the memory number.
3 Press the ENTER key. The data is saved and save mode is canceled.
(To not save the data, press the SAVE key to cancel save mode.)
Tip
To conrm saved data, press the LOAD key.
After pressing the LOAD key, use the ▲▼
key to select the memory number
to conrm. The save date and time appear and then the data appears.
6.1.2 Replacing and Saving
Follow the procedure below to replace saved data (a memory number).
1 Press the SAVE key. (Save mode)
MEM No. (memory number) appears.
2 Use the ▲ ▼ key to select a memory number.
3 Press the ENTER key. If data exists for the memory number,
a replace conrmation indication appears.
4 Press the ENTER key again to replace the data.
The instrument returns to normal (source/measurement) mode.
(To not replace the data, press the SAVE key to cancel the operation.)
Replace confirmation indication
6.1 Data Memory Items