EasyManua.ls Logo

YOKOGAWA CA150 - Page 121

YOKOGAWA CA150
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specications
11-3
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
100 mV
DCV
source
DCmA
source
mA SINK
Resistance
source
RTD output
TC output
1 V
10 V
30 V
20 mA
20 mA
SINK
500 Ω
5 kΩ
50 kΩ
PT 100
JPT 100
K
E
J
N
T
L
U
0 to ±110.000 mV
0 to ±1.10000 V
0 to ±11.0000 V
0 to ±30.00 V
0 to +22.000 µA
0 to -22.000 µA
0 to 550.00 Ω
0 to 5.5000 kΩ
0 to 55.000 kΩ
- 200.0 to 850.0 °C
- 200.0 to 500.0 °C
-200.0 to 1372.0 °C
-200.0 to 1000.0 °C
-200.0 to 1200.0 °C
-200.0 to 400.0 °C
-200.0 to 1300.0 °C
-200.0 to 900.0 °C
-200.0 to 400.0 °C
±(0.04% +25 µV)
±(0.035% +0.1 mV)
±(0.035% +1 mV)
±(0.035% +20 mV)
±(0.04% +7 µA)
±(0.045% +10 µA)
±(0.035% +0.5 Ω)
±(0.065% +5 Ω)
±(0.12% +150 Ω)
±(0.075% +0.6 °C)
±(0.03% +0.8 °C)
However, -200.0 to -100.0 °C
±(0.03% +1.0 °C)
±(0.03% +0.8 °C)
However, -200.0 to -100.0 °C
±(0.03% +1.0 °C)
±(0.03% +0.8 °C)
However, -200.0 to -100.0 °C
±(0.03% +1.0 °C)
±(0.03% +0.9 °C)
±(0.03% +1.0 °C)
However, -200.0 to -100.0
°C
±(0.03% +1.7 °C)
±(0.03% +0.8 °C)
±(0.03% +0.6 °C)
However, -200.0 to 0 °C
±(0.03% +1.2 °C)
R 0 to 1768 °C
±(0.03% +2 °C)
However, 0 to 100 °C
±(0.03% +3.5 °C)
S 0 to 1768 °C
±(0.03% +2 °C)
However, 0 to 100 °C
±(0.03% +3.5 °C)
B 600 to 1820 °C
±(0.03% +2 °C)
However, 600 to 1000 °C
±(0.03% +2.8 °C)
Output resistance:
Approx. 6.5 Ω
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Output resistance:
Approx. 30 mΩ
Maximum output: 10 mA
Maximum load: 24 V
External power supply:
5 to 28 V
Excitation current:
1 to 5 mA*2 or maximum
output: 2V
Excitation current:
0.1 to 0.5 mA or
maximum output: 2V
Excitation current:
0.01 to 0.1 mA or
maximum output: 2V
Excitation current:
1 to 5 mA*2
*3 TC source accuracy
does not include RJ
sensor accuracy.
RJ Sensor Specification
Measurement range:
-10 to 50 °C
Accuracy
(when combined with
main unit)
18 to 28 °C: ±0.5 °C
Other than 18 to 28 °C:
±1 °C
1 µV
Range Source Range Accuracy Remarks
Setting
Resolution
10 µV
0.1 mV
10 mV
1 µA
1 µA
0.01 Ω
0.1 Ω
1 Ω
0.1 °C
0.1 °C
1 °C
*1
*1
*3
Source during Charging
Accuracy = ± (% of setting + µV, mV, µA, , °C) at 23±5°C
Temperature coefcient: Add the accuracy in normal state ×(1/10)°C.
11. Specications

Table of Contents

Related product manuals