Page #94 background image
Communication Function
8-3
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
8.2 Communication Command List
C
ommand Description
When
Normal
When
Setting
OB Queries the battery charge state O X
OD Outputs the measurement values O O
OE Outputs the error information O O
OM Requests sending of memory data O X
OR Queries whether an external RJ sensor is connected O X
OS Outputs the setting information (conditions) O X
PU Sets and queries display of PULSE (source) O X
SD Sets and queries the source values O X
SF Sets and queries the source function O X
SO Starts/stops and queries source O O
SR Sets and queries the source range O O
SY Switches and queries normal mode and calibration mode O O
TE Sets and queries TC and RTD (source) display O X
TT Sets and queries the international temperature standard X O
UP Increases the m (th) digit of the source value by 1 digit O O
VO Sets and queries start/stop of 24 V DC (LOOP) power supply O X