Pannoramic 250 1.15 User’s Guide 5 Troubleshooting and Maintenance
Focus range setting is incorrect
Change focus range setting.
All individual images show a shading effect
Algorithm failed to find empty image fields
Capture a reference image with a blank slide (see section 4.3.2 Scan Options).
5.2.1 Fluorescent Applications (optional)
The shutter of the HXP 120 cannot be triggered, in spite of trigger connection through AxioCam
MRm Rev3.
HXP 120: Shutter is permanently open
Activate shutter remote control with the button at the front of the HXP 120.
The switching of the Lumencor cannot be triggered, in spite of trigger connection through the
camera.
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
HXP 120: Shutter is closed
The opening of the iris contained in the lamp module for fluorescent applications
can be varied by means of the wheel located at the front side of the module. Open
the iris completely.
HXP 120: Intensity is too low
Increase intensity with the rotary button at the front of the HXP 120.
HXP 120 is switched off
Switch light source on.
When using Lumencor
Contact 3DHISTECH Service for support.
March 13, 2012 – Rev. 1 3DHISTECH Ltd.