5-6 Configuring the Analyzer
3. Select Species 1 to create a reference range for a new, unnamed
species, Control 1 to create a range for a new control level, or
Select to create a reference range for a predefined species.
■ To create a reference range for a new, unnamed species:
a. Press Species 1.
b. Set minimum and maximum values for each of the
analytes to be included in the range:
i. Select the analyte to adjust. If needed, use the
up (S) and down (T) arrow keys to scroll
through the list.
ii. Use the controls to set upper and lower limits
for the analyte.
❑ Use the up (S) and down (T) arrow
keys to adjust the values.
❑ Press Clear to set both limits to zero.
❑ Press Default to use the factory default
values for the range.
Note: Entering zero for both values suppresses printing
ranges for this analyte in results (blanks appear
instead), and also suppresses checking the recovered value against the
range limits.
c. Press Save to store the changes.
Add Species
Species 1
Control 1
Select
Back Home
Species 1 Ranges
HomeBack
ALB 0.0 - 0.0
g/dL
ALP
ALT
AMY
AST
0 - 0
0 - 0
0 - 0
0 - 0
U/L
U/L
U/L
U/L
ALB Species 1 Range [g/dL]
SaveCancel
0.0 0.0
Lower
Upper
Clear Default