EasyManua.ls Logo

Abaxis VetScan vs2 - Page 55

Abaxis VetScan vs2
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuring the Analyzer 5-7
To create a new control:
a. Press Control 1.
b. Set minimum and maximum values for each of the
analytes to be included in the range:
i. Select the analyte to adjust from the list. If
needed, use the up (S) and down (T) arrow
keys to view the entire list.
ii. Use the controls to set upper and lower limits
for the analyte.
Use the up (S) and down (T) arrow
keys to adjust the values.
Press Clear to set both limits to zero.
Press Default to use the factory default
values for the range.
Note: Entering zero for both values suppresses printing
ranges for this analyte in results (blanks appear instead), and also sup-
presses checking the recovered value against the range limits.
c. Press Save to store the changes.
To create a reference range for a particular pre-defined species label:
a. Press Select.
b. Select the animal type from those listed. Use the
up (S) and down (T) arrow keys to scroll
through the list.
ALB Control 1 Limit [g/dL]
SaveCancel
0.0 0.0
Lower
Upper
Clear Default

Table of Contents

Related product manuals