7-4 Calibration and Quality Control
1. In the Home screen, press ANALYZE to open the rotor drawer.
2. Insert the control sample into the reagent rotor (see “Running Tests” on page 4-5), and
place the rotor into the drawer.
3. Press CLOSE to close the drawer and start analysis.
4. Enter the Operator ID using the touchscreen, then press
Done.
The operator ID is a number of up to 14 digits that is assigned
by the facility.
The right arrow key ( X) functions as a dash (–), and the left
arrow key (W ) functions as a backspace.
5. Select the control type to use.
Use the up (S) and down (T) arrow keys as needed to scroll
through the list.
6. Enter an ID number for the sample (up to 14 characters), then
press Done.
The right arrow key ( X) functions as a dash (–), and the left
arrow key (W ) functions as a backspace.
Enter Operator ID
DoneCancel
12345678901234
21 3
4
7
5 6
0
98
Select Type
Dog
Cat
Rabbit
Ferret
Control 1
CancelBack