104 15 Sterilization / cleaning / disinfection
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
2. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one
minute.
3. Allow to dry in the air.
Option B: 10% bleach (Clorox EPA REG. NO.: 5813-1)
1. Prepare a 10% bleach solution using purified water.
Use a clean lint-free cloth dampened with the bleach solution.
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE: The treated surface must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one
minute.
4. Allow to dry in the air.
Option C: 5% H
2
O
2
, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H
2
O
2
solution using purified water.
Use a clean lint-free cloth dampened with the H
2
O
2
solution
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE: The treated surface must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one
minute.
4. Allow to dry in the air.
15.2 Accessories
15.2.1 Flow sensor cable
The flow sensor cable must be prepared after each patient treatment.
WARNING:
The flow sensor cable must NOT be sterilized under any circumstances.
15.2.1.1 Preparation for disinfection
1. Switch OFF the fabian before cleaning and unplug at the mains in order to prevent
damage due to penetration by liquid.
2. Remove the Flow sensor from the flow sensor cable.