EasyManua.ls Logo

ACUTRONIC fabian +nCPAP evolution - O2 Therapy mode

ACUTRONIC fabian +nCPAP evolution
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Ventilation modes 83
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
11.8 O
2
Therapy mode
11.8.1 (high and low flow oxygen therapy) HFNC
O
2
Therapy is an option which allows use of a Continuous Flow of blended gas, between 0 to 15
L/min in NEO and 0 to 30 L/MIN in PED mode. Nasal Cannulas of various makes like F&P, Atom or
similar can be used. There are no Alarm functions active in this mode, except for the set FiO
2
NOTE:
This mode can also be used to put the Ventilator in Standby mode. By setting a Flow of 4
L/min, the Humidifier dual servo temperature controls remain active, so there is no need
to switch it OFF in case of short-term Standby mode.
There are no patient alarms active in this mode.
11.9 Ventilation Additives
To optimize ventilation, the following additives can be combined with the selected mode.
See Section 10: Ventilation parameters
11.9.1 Volume limit
The Volume limit function automatically switches to expiration when the default Tidal Volume is
reached.
When changing the ventilation pattern or adjusting the volume limit, the Tidal Volume is
automatically limited.
The display will show the message: Volume Limited.
In the event of Flow Measurement Outage (for example: changing the sensor)
V-limit will automatically be deactivated.
After Flow Measurement is available againV-limitis automatically
reactivated.

Table of Contents

Related product manuals