11.10.2 Nebulizing medications (optional)
WARNING: Flow Sensor wires are HOT.
• If the Flow Sensor remains in the breathing system for extended periods
during Nebulizing without being cleaned, deposits can form from the
medication aerosols and impair Flow Measurement.
• In worst case scenarios these deposits could ignite.
• Disconnecting the plug from the Flow Sensor does NOT prevent the risk
of ignition from medication deposits.
• Always remove the Flow Sensor before Nebulizing Medication.
11.10.3 O
2
Flush / preoxygenation
Press the O
2
Flush button to trigger an O
2
Flush.
Short-term Oxygen Flush with an increased O
2
concentration is permissible
in all Ventilation modes. This will automatically end after a maximum of 2
minutes.
(also see defaults in menu: Configuration -> Ventilation -> Flush Duration)
Press the O
2
Flush button in a Ventilation mode to initiate the O
2
Flush.
The Flush Concentration can now be set with the Rotary push knob.
Pressing the O
2
Flush button again will end the Flush immediately.