11.10.2 Nebulizing medications (optional)
WARNING: Flow Sensor wires are HOT.
If the Flow Sensor remains in the breathing system for extended periods
during Nebulizing without being cleaned, deposits can form from the
medication aerosols and impair Flow Measurement.
In worst case scenarios these deposits could ignite.
Disconnecting the plug from the Flow Sensor does NOT prevent the risk
of ignition from medication deposits.
Always remove the Flow Sensor before Nebulizing Medication.
11.10.3 O
2
Flush / preoxygenation
Press the O
2
Flush button to trigger an O
2
Flush.
Short-term Oxygen Flush with an increased O
2
concentration is permissible
in all Ventilation modes. This will automatically end after a maximum of 2
minutes.
(also see defaults in menu: Configuration -> Ventilation -> Flush Duration)
Press the O
2
Flush button in a Ventilation mode to initiate the O
2
Flush.
The Flush Concentration can now be set with the Rotary push knob.
Pressing the O
2
Flush button again will end the Flush immediately.