38 5 System operation
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
5.4 Device check
Always perform a Device Check before each ventilator use or after changing the
patient circuit.
The fabian ventilator does not automatically compensate gas fraction
measurements for changes in barometric pressure during use. The oxygen sensor is
calibrated in the hospital prior to use and this calibration remains in effect. This
calibration process is part of the normal pre-use checks for the ventilator. The pre-
use calibration should not be skipped if the ventilator has been moved to a
location with significant change in altitude.
What
How
Target
Gas Supply
Attach high pressure Air and Oxygen
supply hoses to the inlets on the rear
panel of the ventilator.
Connect hoses with the
corresponding wall outlets.
Air and Oxygen supply hoses
are correctly connected
Breathing
System
(Dual limb)
Connect the following:
Expiration membrane holder and
expiration membrane
Patient breathing circuit tubes
Water traps (if needed)
Respiratory humidifier and tube
heating
Flow sensor
Test lung
Expiration membrane
holder and membrane
correctly installed.
Patient circuit assembled
correctly according to
manufacturers instructions
Breathing
System
(Single limb)
Connect the following:
Patient breathing circuit tubes
Water traps (if needed)
Respiratory Humidifier and tube
heating
Patient circuit assembled
correctly according to
manufacturers instructions
Switch ON
test
Switch ON the fabian ventilator.
Perform the acoustic audible alarm
test.
Confirm that the audible tone can be
heard
Alarm tone is
audible and alarm
lamp flashes Red
during switch-ON
test.
Self-test successful.
Calibration
Calibrate Flow sensor.
Calibrate O
2
sensor (occurs automatically
after leaving Calibration menu).
Calibration successful

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals