6 Configurations menu 51
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
6.4 Ventilation parameter settings
Figure 6-9: Ventilation parameter settings menu
O2 flush duration (Preoxygenation) configuration:
(OFF / 30 / 60 / 120 sec.)
21 Vol% / 21 and 100 Vol.%. In cases where oxygen supply source
does not provide 100% oxygen, this should be set to 21% only to
avoid calibration at false value.
Set either Off, Low, Middle or High level of leak compensation (0;
up to 10, 20, 30%) in nCPAP and DuoPAP modes.
Selection of pressure unit (mbar / cmH
2
O).
Selection of the nCPAP system used (Infant Flow (Infant Flow or
Inspire), Infant Flow LP, Medijet®).
Maximum time
for Manual
breath
Maximum Inspiration time for Manual breath (2 to 30 seconds).
optional: The “Hospital Settings” button is only active if it is
enabled in service mode.
Resets the device to the hospital defaults.
optional: The “factory defaults” button is only active if it is
enabled in service mode.
Resets the device to the factory defaults.