52 6 Configurations menu
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Flow Trigger / Volume Trigger / Pressure Trigger. Trigger mode
configuration for assisted breathing
After visual reset of alarm, the reactivation of alarm is delayed for
the set time period. This avoids an immediate re-alarm in case
Operational parameters have not yet stabilized.
Transport mode should be selected when the fabian is used for
intra-hospital transport. The maximum inspiratory pressure
value is restricted to 60 mbar to prolong battery run time.
Setting of the flow during the expiratory phase of the breathing
cycle
I-time Display Configuration. Setting I time or frequency
(Frequency - I-time / I-time / E-time).
PSV flow termination criteria
Select Ppsv parameter to be absolute value or above PEEP
Parameter.
Choice of Volume Limit or Volume Guarantee setting
Select if tidal volume measurement should be updated for every
breath.
WARNING:
Only Reset to hospital and factory settings when the device is NOT connected to patients.