78 11 Ventilation modes
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
11.2 SIPPV
During Synchronized Intermittent
Positive Pressure Ventilation (SIPPV),
each spontaneous patient inspiratory
effort triggers a mechanical breath of the
ventilator according to the ventilator
parameters set for Inspiratory Period and
Pressure.
The number of breaths per minute
supported by the ventilator is controlled
by the patient. In this mode the patient
breathing frequency must be closely
monitored to prevent hyperventilation.
Set the Alarm for high Frequency
according clinical guidelines.
Set following parameters:
Eflow.
Frequency (Rate)
Iflow
I-Time
PEEP
Pinsp
Trigger Sensitivity
To prevent Auto-Triggering, another
breath cannot be triggered for 180 ms
following the end of the preceding
breath.
See Section 10: Ventilation parameters
See Section 11.9: Ventilation Additives